The total length of the NMR tube must NOT exceed 9 inches
Inserting longer tubes is possible with bypassing the SampleCase. See Manual Insertion section below.
Ceramic spinners may be used in the automatic mode. To ensure reliable ejection, you must have VTU turned ON with a VT gas flow set to 400 LPH before inserting/ejecting. You may consult Manual Insertion on adjustment of the VT gas flow.
SIDE NOTE: Alternatively, you may use the Bruker 500 MHz system in McCourtney, which does NOT have a limitation on the tube length. However, ejecting ceramic spinners with heavy tubes in McCourtney 500 may still be a challenge. You may consult Manual Insertion on adjustment of the VT gas flow to help ejection.
The Bruker SampleCase sample changer is installed at Bruker 400 MHz and 800 MHz systems in Stepan Chemistry Hall and Bruker 400 MHz automated system in McCourtney Hall. This overview shows you main parts of the SampleCase system
NOTE: You MUST always set the tube depth with a depth gauge!
If your tube is set deeper than required, it will be broken by a changer and may damage the probe!
If your tube is does NOT exceed 9 inches (including a cap), it WILL safely fit in the SampleCase.
This video shows how SampleCase detects excessively long tubesĀ and how to retrieve on "offender" tube.
If in doubt, after you set the tube to a correct depth in a spinner, you may check the height with a ruler. The changer will stop if the tube exceeds 97.5 mm above the spinner:
Manual Sample Insertion (bypassing the SampleCase)
IMPORTANT: The manual Insertion in SampleCase is allowed on Bruker 400 MHz system in Stepan Chemistry Hall but not Bruker 800 MHz or automatic 400 MHz systems.
For using the manual insertion mode, you have to request an authorization from the NMR staff.
Long tubes exceeding 9 inches may be inserted manually bypassing the changer. Ceramic spinners may also be difficult to eject in automatic mode, therefore, should be inserted manualy. The video shows you the process:
IMPORTANT: Never touch the shuttle with your hand!
Summary
To insert your sample manually, do this:
Step 1. Eject current sample
ro off
lock off
sx ej
Step 2. Switch to manual insert mode
Turn the blue valve to a horizontal position.
Open Sample Transporter Control screen:
Switch Lift Mode to Standard BSMS Lift
Click Set
Increase VT gas flow to use with J Young tubes and/or Ceramic spinners:
Enter edte screen
Set Target Temperature to the value currently displayed
Set Target Gas Flow to 600 LPH
NOTE: Pay attention not to change the Standby Gas Flow!
NOTE 2:
Light samples may need lower Target Gas Flow rate (500-400 LPH), otherwise they may not insert (remain seated in the probe)!
Heavy samples may need higher Target Gas Flow rate (800+ LPH) for the instrument to be able to eject them!
Turn VTU system ON.
Step 3. Insert your sample manually
Type ej
Place your tube in the bore
Type ij
Step 4. Returning to SampleCase configuration
After your experiments are finished, you MUST return the instrument to its standard SampleCase configuration.
To do this:
Eject your sample:
ro off
lock off
ej
Pick up your sample and leave the bore empty.
Type ij
Open Sample Transporter Control screen.
Switch Lift Mode to Sample Transporter.
Click Set
Turn Blue Valve to a vertical position.
Return to a standard VT gas flow:
Enter edte screen
Turn VTU system OFF.
Set Target Gas Flow to 200 LPH NOTE: Pay attention not to change Standby Gas Flow!